LP 2000 Bench top Turbidity Meter
- LED Light Source LP 2000 is the
only bench turbidity meter on the market with this superior feature.
- Calibration Date
Storage An unique GLP (Good Laboratory Practice) feature.
- Accuracy Unsurpassed
accuracy with ±0.5 FTU through an extensive range.
|
LP 2000 is the lab operators dream instrument! It measures turbidity from
0 to 50 FTU* with 0.01 sensitivity and then reads all the way from 50 to
1000 FTU with a 1 FTU resolution. Calibration is automatic, quick and simple.
Two calibration standards with fixed values of 0 and 10 FTU are
automatically recognized by the microprocessor. The standards are
supplied with the instruments and are AMCO-AEPA-1 instead of formazine
since they are more stable, non-toxic and can be reused for up to a year
without contamination.
LP 2000 uses the infrared beam
and the 90° sensor method to measure turbidity since it is less sensitive
to colour as opposed to the filter system. LP 2000 is the only turbidity
meter equipped with an infrared LED as source of light. The infrared LED
has a long life, unlike standard lamps, and more importantly, it
maintains constant emission for the entire life of the instrument. The
wavelength peaks at 890 nm, which ensures maximum intensity of diffused light
even in samples with low turbidity values, and also reduces the
interference from any colours. A large LCD display shows read-outs in
clear digits.
The GLP
feature automatically stores and retrieves the last calibration date. At
the touch of a key, the last calibration data is displayed allowing the
user to ensure accurate calibration with every test.
LP 2000 bench turbidity meter
is supplied with measurement cuvet and cap, HI 710005 or HI 710006 12VDC
adapter, HI 93703-0 and HI 93703-10 calibration solutions and instruction
manual.
|
Specifications:
|
LP 2000
|
Range
|
0.00 to 50.00 FTU / 50 to 1000 FTU*
|
Resolution
|
0.01 / 1 FTU
|
Accuracy
|
±0.5 FTU or ±5% (whichever greater)
|
Typical EMC Deviation
|
0.2 FTU
|
Light Source/Life
|
High Emission Infrared LED / Life of the instrument
|
Light Detector
|
Silicon Photocell
|
Power Supply
|
12VDC through HI 710005 or HI 710006 (included)
|
Environment
|
32 to 122°F (0 to 50°C); RH
95%
|
Dimensions
|
9.1 x 6.7 x 2.7" (230 x 170 x 70 mm)
|
Weight
|
1.3 lb (600 g)
|
|
|
|
1 FTU (Formazine Turbidity Unit) = 1 NTU
(Nephelometric Turbidity Unit)
Back to
Top
|

|
C 114 Turbidity and Chlorine Measurement
- USEPA Compliance - C 114
performs turbidity measurements using the nephelometric principle
according to USEPA's 180.1 method and the Standard Method 2130B.
- Good Laboratory
Practice (GLP) - With the press of a button, the last calibration
information is retrieved from memory and displayed to assure
accurate measurements.
- Logging
Capability - This powerful feature allows you to store information
from up to 25 samples including the time and date of each
measurement.
|
With the hand
held C 114, three important parameters in water quality analysis are
at your fingertips. Free and total chlorine measurements and EPA compliant
turbidity measurements are performed quickly and reliably with this
lab-grade microprocessor-based field instrument. C 114 is
particularly useful for potable water measurements and municipalities.
The rugged C 114 will log data for up to
25 measurements along with time and date information for retrieval at a
later date. This instrument also offers Good Laboratory Practice (GLP)
features which allow the user to verify the calibration status of the
meter. With C 114 measurements can be performed using either
concentration or relative absorbance of the sample depending on the
requirements of the application. C 114 can be calibrated using
the factory pre-programmed calibration points or can be custom calibrated
when in colorimetric mode. This rugged instrument has been engineered to
offer all the sophistication and reliability of benchtop laboratory
instrumentation in the size of a hand-held meter.
C 114 portable
turbidity/chlorine meter is supplied with measurement cuvet and cap, 4 x
1.5V AA batteries and instruction manual.
Specifications:
|
|
C 114
|
Range
|
Turbidity
|
0.00 to 9.99
NTU & 10.0 to 50.0 NTU*
|
Free Cl2
|
0.00 to 2.50
mg/L
|
Total Cl2
|
0.00 to 3.50
mg/L
|
Resolution
|
Turbidity
|
0.01 & 0.1
NTU
|
Cl2
|
0.01 mg/L
|
Accuracy
|
Turbidity
|
±0.5 NTU or ±5%
of reading (whichever greater)
|
Cl2
|
±0.03 mg/L ±3%
of reading
|
Light Source/Life
|
Pure Green LED
/ Life of the instrument
|
Light Detector
|
Two Silicon
Photocells
|
Battery Type/Life
|
4 x 1.5V AA
batteries 60 hours or 1000 measurements Auto-shut off user-selectable
from 10, 20, 30, 40, 50 or 60 minutes
|
Environment
|
32 to 122°F (0
to 50°C); RH 95%
|
Dimensions
|
8.7 x 3.2 x
2.6" (220 x 82 x 66 mm)
|
Weight
|
18 oz. (510 g)
|
1 FTU (Formazine Turbidity Unit) = 1 NTU
(Nephelometric Turbidity Unit)
Back to
Top

|
HI 93703 Portable Turbidity Meter Complying with ISO
7027
- LED Light Source HI 93703 is the
only portable turbidity meter on the market with this superior
feature.
- Calibration Date
Storage The unique GLP (Good Laboratory Practice) feature.
- Supplied with
Starter Kit HI 93703 portable turbidity meter is supplied with
measurement cuvet & cap, 4 x 1.5V AA batteries and instruction
manual. HI 93703KIT is comprised of HI 93703 complete with HI 731313
maintenance kit
- HI 93703C (with
case)
|
The I 93703 portable microprocessor-based turbidity
meter provides lab-grade accuracy even in the field. With 4 keys and
weighing only 350 grams, HI 93703 measures turbidity from 0 to 1000 FTU (NTU)*.
For best field accuracy, it measures from 0 to 50 FTU in steps of 1/100th
of FTU. This is 10 times better than comparable models on the market that
measure 10 to 50 FTU only in decimal points! The meter meets the ISO 7027
requirements.
HI
93703 is the only turbidity meter equipped with an infrared LED as
source of light. Unlike standard lamps, the infrared LED has a long life.
More significantly, it maintains constant emission for the entire life of
the instrument. The wavelength peaks at 890 nm, which provides the
required intensity of diffused light even in samples with low turbidity
values, and also reduces the interference from any colors.
The meter is
very easy to use. All operations can be carried out with only four keys
and troubleshooting functions can be performed with error code on LCD.
Moreover, the meter's versatility and durability ensure low maintenance.
GLP circuitry
automatically stores and retrieves the last calibration data. At the
touch of a key, the last calibration data together with time and date are
displayed allowing the user to maintain an accurate calibration
scheduleAuto shut-off turns the meter off after 5 minutes of non-use to
save batteries. HI 93703 is available at the most competitive price on
the market.
|
Specifications:
|
HI 93703
|
Range
|
0.00 to 50.00
FTU / 50 to 1000 FTU*
|
Resolution
|
0.01 / 1 FTU
|
Accuracy
|
±0.5 FTU or ±5%
(whichever greater)
|
Typical EMC Deviation
|
±2% Full Scale
|
Light Source/Life
|
High Emission
Infrared LED / Life of the instrument
|
Light Detector
|
Silicon
Photocell
|
Battery Type/Life
|
4 x 1.5V AA
batteries 60 hours or 900 measurements Auto-shut off after 5 minutes
|
Environment
|
32 to 122°F (0
to 50°C); RH 95%
|
Dimensions
|
8.7 x 3.2 x
2.6" (220 x 82 x 66 mm)
|
Weight
|
18 oz. (510 g)
|
1 FTU (Formazine Turbidity Unit) = 1 NTU
(Nephelometric Turbidity Unit)
Back to
Top
|

|
C102 Turbidity, Chlorine, pH, Bromine, Iron, Iodine
& Cyanuric Acid
- USEPA Compliance In the
turbidimetric mode, C 102 uses the nephelometric principle according
to USEPA's 180.1 method and the Standard Method 2130B.
- Good Laboratory
Practice (GLP) With a built-in clock, you can check the last calibration
date at a touch of a button to guarantee best performance for your
meter.
- Custom
Calibration Points Hi-tech electronics allow the operator to
calibrate the meter at a selectable point in from 0.00 to 50.00
NTU*.
- Logging
Capability C 102 makes it possible to log and retrieve up to 25
different samples.
- Economical C 102 is the most
competitively price meter around.
|
The
most important parameters needed for water analysis, specially in drinking
water, are now available through a portable meter. You can obtain lab
performance in the field by using C 102. It not only
measures turbidity, but also pH, Total and Free Chlorine, Bromine,
Iodine, Low Iron and Cyanuric Acid (CYS). All this in the palm of your
hand and in seconds!
Measurements
are made quickly and repeatedly through a sophisticated, yet easy-to-use
microprocessor. In colorimetric mode, users can select between factory
preprogrammed calibration or calibrating the meter on their own, and
measure either concentration or relative absorbance of the sample. Up to
25 measured samples can be stored in memory together with time and date.
Miniaturization of the electronics has made it possible to offer
unsurpassed accuracy and quality in a portable unit weighing just one
pound!
C 102
portable
water analysis meter is supplied with measurement cuvet & cap, 4 x
1.5V AA Batteries and instruction manual.
|
Specifications:
|
|
C 102
|
Range
|
Turbidity
|
0.00 to 9.99
NTU & 0.0 to 50.0 NTU*
|
|
Bromine
|
0.00 to 8.00
mg/L
|
|
Free Chlorine
|
0.00 to 2.50
mg/L
|
|
Total Chlorine
|
0.00 to 3.50
mg/L
|
|
Cyanuric Acid
|
0 to 80 mg/L
|
|
Iodine
|
0.0 to 12.5
mg/L
|
|
Iron
|
0.00 to 1.00
mg/L
|
|
pH
|
5.9 to 8.0 pH
|
Resolution
|
Turbidity
|
0.01 & 0.1
NTU
|
|
Bromine
|
0.01 mg/L
|
|
Free Chlorine
|
0.01 mg/L
|
|
Total Chlorine
|
0.01 mg/L
|
|
Cyanuric Acid
|
1 mg/L
|
|
Iodine
|
0.1 mg/L
|
|
Iron
|
0.01 mg/L
|
|
pH
|
0.1 pH
|
Accuracy
|
Turbidity
|
±0.5% NTU or
±5% of reading (whichever greater)
|
|
Bromine
|
±0.2 mg/L ±3%
of reading
|
|
Free Chlorine
|
±0.02 mg/L ±3%
of reading
|
|
Total Chlorine
|
±0.02 mg/L ±3%
of reading
|
|
Cyanuric Acid
|
±1 mg/L ±15% of
reading
|
|
Iodine
|
±0.1 mg/L ±5%
of reading
|
|
Iron
|
±0.02 mg/L ±3%
of reading
|
|
pH
|
±0.1 pH
|
Typical EMC Deviation
|
Turbidity
|
±2% Full Scale
|
Bromine
|
±0.01 mg/L
|
|
Free Chlorine
|
±0.01 mg/L
|
|
Total Chlorine
|
±0.01 mg/L
|
|
Cyanuric Acid
|
±1 mg/L
|
|
Iodine
|
±0.1 mg/L
|
|
Iron
|
±0.01 mg/L
|
|
pH
|
±0.1 pH
|
Light Source/Life
|
|
Pure Green LED
/ Life of the instrument
|
Light Detector
|
|
Two Silicon
Photocells
|
Battery Type/Life
|
|
4 x 1.5V AA batteries
60 hours or 1,000 measurements / Auto-shut off user-selectable from 10,
20, 30, 40, 50 or 60 minutes
|
Environment
|
|
32 to 122°F (0
to 50°C); RH 95%
|
Dimensions
|
|
8.7 x 3.3 x
1.8" (220 x 83 x 46 mm)
|
Weight
|
|
18 oz. (510 g)
|
1 FTU (Formazine Turbidity Unit) = 1 NTU
(Nephelometric Turbidity Unit)
Back to Top
|