K88600
Portable Octane & Cetane Analyzer
The analyzer measures octane number via near-infrared (NIR) transmission
spectroscopy utilizing 14 near-infrared emitting diodes with narrow bandpass
filters, a silicon detector system, and a fully integrated microprocessor.
Simple octane number determination requires three easy steps: sampling a
background signal, acquiring two absorption spectra of the gas sample, and
then acquiring a second background signal. Analyzer is pre-calibrated for
unleaded gasoline and ethanol-blended fuels, and can be calibrated for up to
eight additional fuel types.
The analyzer is small, lightweight, and operates on AA" batteries or AC.
Before each reading, the unit standardizes itself to assure accuracy. The
octane number is printed with time and date on the built-in printer. All data
can be downloaded via the RS232 port to an external computer.
Specifications
Accuracy and repeatability equivalent to ASTM approved CFR engines test
methods (ASTM D2699, D2700)
Sample Holder: Sealed, cubical glass container (75mm optical path length)
Sample Volume: 8 Ounces (approx. 225 mL)
Precalibrated for unleaded gasoline & ethanol-blended fuels. (Analyzer
can be calibrated for up to 8 additional fuel types.)
Battery operated (6 AA batteries)
Included Accessories
IBM Compatible Software
RS232 Cable
Aluminum Carrying Case
5 Rolls of paper
3 Sample Holders
Light Cover
6 AA Batteries
6 Sample Holder Labels
Dimensions lxwxh,in. (cm)
13.5x4.5x2.5 (34x11.5x61&Mac218;4)
Net Weight: 12 lbs (5.5kg)
Shipping Information
23x17x8.5 (58.5x43.25x22)
Shipping weight: 25 lbs (11.5kg)
|