K80100
Dual Corrosion Test Apparatus
Consists of two heat-transfer corrosion cells with temperature and pressure
controls for evaluating engine coolant samples in accordance with ASTM D4340.
Maintains aluminum alloy specimens at constant temperature in the presence of
500mL samples. Each corrosion cell includes borosilicate glass test cell,
heat transfer bar, aluminum test specimen with thermocouple port, Type J
thermocouple, 950W band heater, pressure gauge, pressure relief valve,
stainless steel end plates with support rods, fill plug and O-rings, and
acrylic safety shields. Digital temperature controllers maintain specimens at
temperatures of up to 210 ± 1°C throughout the test period. Test cells are
mounted on a finished steel base. Controls are housed in a separate finished
steel cabinet.
Specifications
Conforms to the specifications of: ASTM D4340
Test Capacity: 2 coolant samples
Maximum Temperature: 210 °C
Electrical Requirements
115V 50/60Hz, 17A
220-240V 50/60Hz, 8.5A
Dimensions lxwxh,in.(cm)
Test Unit: 43x27x44 (109x62x112)
Control Cabinet: 10x14x12 (25x36x30)
Net Weight: 190 lbs (86.2kg)
Shipping Information
Shipping Weight: 327 lbs (148.3kg)
Dimensions: 46.7 Cu. ft.
|