Pensky-Martens Closed Cup Flash Point Tester
The automated Pensky-Martens flash point tester accurately determines the
lowest flash point temperature of fuels, lubricating oils, and homogenous
liquids (ASTM D93 A), or liquids containing suspended solids as well as
liquids that tend to form a surface film during testing (ASTM D93 B). Flash point
tests are simply conducted by mounting the flash cup filled with sample into
the test position and selecting a pre- or user-programmed test method. The
automation routines provide accurate test results, even with users
inexperienced in flash point test methods. The flash point test result is
automatically corrected to standard pressure (101.3 kPa). The unit is
equipped with a differential Pt-100 RTD probe designed to duplicate the
response time of a mercury-in-glass thermometer and with multiple sensors for
continually monitoring of instrument function, displaying an error message if
a problem is detected. The performance of the electrical ignitor is
continuously checked, and the user is notified upon the need of replacement
due to either damage or the end of its useful life. When performing a test,
the user is also alerted if the first application of the ignitor results in a
flash or if no flash point is detected at the end of the test program. If a
flash is not detected 30°C above the expected flash point or at 400°C, then
the test is automatically aborted for safety.
Specifications
Conforms to the specifications of:
ASTM D93; IP 34; ISO 2719; DIN EN 22719; NF M 07-019; JIS K2265
Electrical Requirements
115V 50/60Hz, Single Phase
230V 50/60Hz, Single Phase
Dimensions lxwxh,in.(cm)
15.3x9x18 (39x23x46)
Net Weight:
17.5 lbs (8kg)
|