Automatic
Tag Closed Cup Flash Point Tester
The automated Tag Closed Cup flash point tester ensures the accuracy and
precision required according to the ASTM D56 and related test method. The test
sample is heated at a prescribed rate of temperature increase throughout the
standard temperature test range of 0 to 100°C, and the extended temperature
model achieves a range of –30 to 100°C with the use of an external chiller.
The flash point tests are simply conducted by mounting the flash cup filled
with sample into the test position and selecting a pre-programmed test method
or the search mode to determine an approximate flash point. The automation
routines provide accurate test results, even with users inexperienced in
flash point test methods. Ignition by gas flame or electrical ignitor are
included on both models, along with safety cut-off devices. The test results
are automatically corrected to standard pressure (101.3 kPa). The system is
equipped with a differential Pt-100 RTD probe designed to duplicate the
response time of a mercury-in-glass thermometer and with multiple sensors
that continually monitor instrument function, displaying an error message if
a problem is detected.
Specifications
Conforms to the specifications of:
ASTM D56; IP 304
Electrical Requirements
115V 50/60Hz, Single Phase
230V 50/60Hz, Single Phase
Dimensions lxwxh,in.(cm)
15.3x9x18 (39x23x46)
Net Weight:
17.5 lbs (8kg)
|